OBJECTIVES: The purpose of this study is to assess the effectiveness of defibrillation testing (DT) in patients undergoing implantable cardioverter-defibrillator (ICD) insertion. BACKGROUND: Although DT is considered a standard procedure during ICD implantation, its usefulness has not been definitively proven. METHODS: The SAFE-ICD (Safety of Two Strategies of ICD Management at Implantation) study is a prospective observational study designed to evaluate the outcome of 2 strategies: performing defibrillation testing (DT+) versus not performing defibrillation testing (DT-) during de novo ICD implants. No deviation from the centers' current practice was introduced. In all, 2,120 consecutive patients (836 DT+ and 1,284 DT-) age ≥18 years were enrolled at 41 Italian centers from April 2008 to May 2009 and followed up for 24 months until June 2011. The primary endpoint was a composite of severe complications at ICD implant and sudden cardiac death or resuscitation at 2 years. RESULTS: The primary endpoint occurred in 34 patients: 12 intraoperative complications (8 in DT+ group; 4 in DT- group) and 22 during follow-up (10 in DT+ group; 12 in DT- group). Overall, the estimated yearly incidence (95% confidence interval) was DT+ 1.15% (0.73 to 1.83) and DT- 0.68% (0.42 to 1.12). The difference between the 2 groups was negligible: 0.47% per year (-0.15 to 1.10). Mortality from any cause was similar at 2 years (adjusted hazard ratio: 0.97 [0.76 to 1.23], p = 0.80). CONCLUSIONS: In this large cohort of new ICD implants, event rates were similar and extremely low in both groups. These data indicate a limited clinical relevance for DT testing, thus supporting a strategy of omitting DT during an ICD implant. (Safety of Two Strategies of ICD Management at Implantation [SAFE-ICD]; NCT00661037).

Clinical Evaluation of Defibrillation Testing in an Unselected Population of 2,120 Consecutive Patients Undergoing First Implantable Cardioverter-Defibrillator Implant

FAVALE, Stefano;
2012-01-01

Abstract

OBJECTIVES: The purpose of this study is to assess the effectiveness of defibrillation testing (DT) in patients undergoing implantable cardioverter-defibrillator (ICD) insertion. BACKGROUND: Although DT is considered a standard procedure during ICD implantation, its usefulness has not been definitively proven. METHODS: The SAFE-ICD (Safety of Two Strategies of ICD Management at Implantation) study is a prospective observational study designed to evaluate the outcome of 2 strategies: performing defibrillation testing (DT+) versus not performing defibrillation testing (DT-) during de novo ICD implants. No deviation from the centers' current practice was introduced. In all, 2,120 consecutive patients (836 DT+ and 1,284 DT-) age ≥18 years were enrolled at 41 Italian centers from April 2008 to May 2009 and followed up for 24 months until June 2011. The primary endpoint was a composite of severe complications at ICD implant and sudden cardiac death or resuscitation at 2 years. RESULTS: The primary endpoint occurred in 34 patients: 12 intraoperative complications (8 in DT+ group; 4 in DT- group) and 22 during follow-up (10 in DT+ group; 12 in DT- group). Overall, the estimated yearly incidence (95% confidence interval) was DT+ 1.15% (0.73 to 1.83) and DT- 0.68% (0.42 to 1.12). The difference between the 2 groups was negligible: 0.47% per year (-0.15 to 1.10). Mortality from any cause was similar at 2 years (adjusted hazard ratio: 0.97 [0.76 to 1.23], p = 0.80). CONCLUSIONS: In this large cohort of new ICD implants, event rates were similar and extremely low in both groups. These data indicate a limited clinical relevance for DT testing, thus supporting a strategy of omitting DT during an ICD implant. (Safety of Two Strategies of ICD Management at Implantation [SAFE-ICD]; NCT00661037).
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11586/89490
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 62
  • ???jsp.display-item.citation.isi??? 50
social impact