BACKGROUND The HeartLogic algorithm combines multiple implantable cardioverter-defibrillator sensors to identify patients at risk of heart failure (HF) events. We sought to evaluate the risk stratification ability of this algorithm in clinical practice. We also analyzed the alert management strategies adopted in the study group and their association with the occurrence of HF events. METHODS The HeartLogic feature was activated in 366 implantable cardioverter-defibrillator and cardiac resynchronization therapy implantable cardioverter-defibrillator patients at 22 centers. The median follow-up was 11 months [25th-75th percentile: 6-16]. The HeartLogic algorithm calculates a daily HF index and identifies periods IN alert state on the basis of a configurable threshold. RESULTS The HeartLogic index crossed the threshold value 273 times (0.76 alerts/patient-year) in 150 patients. The time IN alert state was 11% of the total observation period. Patients experienced 36 HF hospitalizations, and 8 patients died of HF during the observation period. Thirty-five events were associated with the IN alert state (0.92 events/patient-year versus 0.03 events/patient-year in the OUT of alert state). The hazard ratio in the IN/OUT of alert state comparison was (hazard ratio, 24.53 [95% CI, 8.55-70.38], P<0.001), after adjustment for baseline clinical confounders. Alerts followed by clinical actions were associated with less HF events (hazard ratio, 0.37 [95% CI, 0.14-0.99], P=0.047). No differences in event rates were observed between in-office and remote alert management. CONCLUSIONS This multiparametric algorithm identifies patients during periods of significantly increased risk of HF events. The rate of HF events seemed lower when clinical actions were undertaken in response to alerts. Extra in-office visits did not seem to be required to effectively manage HeartLogic alerts. Registration: URL: ; Unique identifier: NCT02275637.

Multiparametric Implantable Cardioverter-Defibrillator Algorithm for Heart Failure Risk Stratification and Management: An Analysis in Clinical Practice

Vincenzo Ezio Santobuono;F Ammirati;F Guerra;V Amato;A Costa;D Porcelli;
2021-01-01

Abstract

BACKGROUND The HeartLogic algorithm combines multiple implantable cardioverter-defibrillator sensors to identify patients at risk of heart failure (HF) events. We sought to evaluate the risk stratification ability of this algorithm in clinical practice. We also analyzed the alert management strategies adopted in the study group and their association with the occurrence of HF events. METHODS The HeartLogic feature was activated in 366 implantable cardioverter-defibrillator and cardiac resynchronization therapy implantable cardioverter-defibrillator patients at 22 centers. The median follow-up was 11 months [25th-75th percentile: 6-16]. The HeartLogic algorithm calculates a daily HF index and identifies periods IN alert state on the basis of a configurable threshold. RESULTS The HeartLogic index crossed the threshold value 273 times (0.76 alerts/patient-year) in 150 patients. The time IN alert state was 11% of the total observation period. Patients experienced 36 HF hospitalizations, and 8 patients died of HF during the observation period. Thirty-five events were associated with the IN alert state (0.92 events/patient-year versus 0.03 events/patient-year in the OUT of alert state). The hazard ratio in the IN/OUT of alert state comparison was (hazard ratio, 24.53 [95% CI, 8.55-70.38], P<0.001), after adjustment for baseline clinical confounders. Alerts followed by clinical actions were associated with less HF events (hazard ratio, 0.37 [95% CI, 0.14-0.99], P=0.047). No differences in event rates were observed between in-office and remote alert management. CONCLUSIONS This multiparametric algorithm identifies patients during periods of significantly increased risk of HF events. The rate of HF events seemed lower when clinical actions were undertaken in response to alerts. Extra in-office visits did not seem to be required to effectively manage HeartLogic alerts. Registration: URL: ; Unique identifier: NCT02275637.
File in questo prodotto:
File Dimensione Formato  
hhf-14-e008134.pdf

accesso aperto

Descrizione: Article
Tipologia: Documento in Versione Editoriale
Licenza: Creative commons
Dimensione 480.38 kB
Formato Adobe PDF
480.38 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11586/448908
Citazioni
  • ???jsp.display-item.citation.pmc??? 20
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 33
social impact