A photoacoustic detection module based on a gold-plated photoacoustic cell was reported in this manuscript to measure hydrogen sulfide (H2S) gas in sewers. A 1582 nm distributed feedback (DFB) diode laser was employed as the excitation light source of the photoacoustic sensor. Operating pressure within the photoacoustic cell and laser modulation depth were optimized at room temperature, and the long-term stability of the photoacoustic sensor system was analyzed by an Allan-Werle deviation analysis. Experimental results showed that under atmospheric pressure and room temperature conditions, the photoacoustic detection module exhibits a sensitivity of 11.39 mu V/ppm of H2S and can reach a minimum detection limit (1 sigma) of 140 ppb of H2S with an integration time of 1 s. The sensor was tested for in-field measurements by sampling gas in the sewer near the Shanxi University canteen: levels of H2S of 81.5 ppm were measured, below the 100 ppm limit reported by the Chinese sewer bidding document.

Detection of Hydrogen Sulfide in Sewer Using an Erbium-Doped Fiber Amplified Diode Laser and a Gold-Plated Photoacoustic Cell

Giglio, Marilena;Sampaolo, Angelo;Patimisco, Pietro;Spagnolo, Vincenzo;
2022-01-01

Abstract

A photoacoustic detection module based on a gold-plated photoacoustic cell was reported in this manuscript to measure hydrogen sulfide (H2S) gas in sewers. A 1582 nm distributed feedback (DFB) diode laser was employed as the excitation light source of the photoacoustic sensor. Operating pressure within the photoacoustic cell and laser modulation depth were optimized at room temperature, and the long-term stability of the photoacoustic sensor system was analyzed by an Allan-Werle deviation analysis. Experimental results showed that under atmospheric pressure and room temperature conditions, the photoacoustic detection module exhibits a sensitivity of 11.39 mu V/ppm of H2S and can reach a minimum detection limit (1 sigma) of 140 ppb of H2S with an integration time of 1 s. The sensor was tested for in-field measurements by sampling gas in the sewer near the Shanxi University canteen: levels of H2S of 81.5 ppm were measured, below the 100 ppm limit reported by the Chinese sewer bidding document.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11586/424053
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 7
social impact