Quartz-enhanced photoacoustic spectroscopy (QEPAS) with a single-tube acoustic microresonator (AmR) inserted between the prongs of a custom quartz tuning fork (QTF) was developed, investigated, and optimized experimentally. Due to the high acoustic coupling efficiency between the AmR and the QTF, the single-tube on-beam QEPAS spectro-phone configuration improves the detection sensitivity by 2 orders of magnitude compared to a bare QTF. This approach significantly reduces the spectrophone size with respect to the traditional on-beam spectrophone configuration, thereby facilitating the laser beam alignment. A 1(7 normalized noise equivalent absorption coefficient of 1.21 × 10-8 cm-1. W/√Hz was obtained for dry CO2 detection at normal atmospheric pressure.
Single-tube on-beam quartz-enhanced photoacoustic spectroscopy
Sampaolo A.;Patimisco P.;Spagnolo V.;
2016-01-01
Abstract
Quartz-enhanced photoacoustic spectroscopy (QEPAS) with a single-tube acoustic microresonator (AmR) inserted between the prongs of a custom quartz tuning fork (QTF) was developed, investigated, and optimized experimentally. Due to the high acoustic coupling efficiency between the AmR and the QTF, the single-tube on-beam QEPAS spectro-phone configuration improves the detection sensitivity by 2 orders of magnitude compared to a bare QTF. This approach significantly reduces the spectrophone size with respect to the traditional on-beam spectrophone configuration, thereby facilitating the laser beam alignment. A 1(7 normalized noise equivalent absorption coefficient of 1.21 × 10-8 cm-1. W/√Hz was obtained for dry CO2 detection at normal atmospheric pressure.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.