A quartz-enhanced photoacoustic spectroscopy sensor system was developed for the sensitive detection of hydrogen peroxide (H2O2) using its absorption transitions in the v6 fundamental band at ∼7.73 μm. The recent availability of distributed-feedback quantum cascade lasers provides convenient access to a strong H2O2 absorption line located at 1295.55 cm-1. Sensor calibration was performed by means of a water bubbler that generated titrated average H2O2 vapor concentrations. A minimum detection limit of 12 parts per billion (ppb) corresponding to a normalized noise equivalent absorption coefficient of 4.6 × 10-9 cm-1W/Hz1/2 was achieved with an averaging time of 100 s.
Hydrogen peroxide detection with quartz-enhanced photoacoustic spectroscopy using a distributed-feedback quantum cascade laser
Patimisco P.;Spagnolo V.;
2014-01-01
Abstract
A quartz-enhanced photoacoustic spectroscopy sensor system was developed for the sensitive detection of hydrogen peroxide (H2O2) using its absorption transitions in the v6 fundamental band at ∼7.73 μm. The recent availability of distributed-feedback quantum cascade lasers provides convenient access to a strong H2O2 absorption line located at 1295.55 cm-1. Sensor calibration was performed by means of a water bubbler that generated titrated average H2O2 vapor concentrations. A minimum detection limit of 12 parts per billion (ppb) corresponding to a normalized noise equivalent absorption coefficient of 4.6 × 10-9 cm-1W/Hz1/2 was achieved with an averaging time of 100 s.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.