The design and realization of two highly sensitive and easily interchangeable spectrophones based on custom quartz tuning forks, with a rectangular (S1) or T-shaped (S 2 ) prongs geometry, is reported. The two spectrophones have been implemented in a QEPAS sensor for ethylene detection, employing a DFB-QCL emitting at 10.337 μm with an optical power of 74.2 mW. A comparison between their performances showed a signal-to-noise ratio 3.4 times higher when implementing the S 2 spectrophone. For the S 2 -based sensor, a linear dependence of the QEPAS signal on ethylene concentration was demonstrated in the 5 ppm-100 ppm range. For a 10 s lock-in integration time, an ethylene minimum detection limit of 10 ppb was calculated.
Quartz-enhanced photoacoustic sensor for ethylene detection implementing optimized custom tuning fork-based spectrophone
Giglio M.;Elefante A.;Patimisco P.;Sampaolo A.;Sgobba F.;Spagnolo V.
2019-01-01
Abstract
The design and realization of two highly sensitive and easily interchangeable spectrophones based on custom quartz tuning forks, with a rectangular (S1) or T-shaped (S 2 ) prongs geometry, is reported. The two spectrophones have been implemented in a QEPAS sensor for ethylene detection, employing a DFB-QCL emitting at 10.337 μm with an optical power of 74.2 mW. A comparison between their performances showed a signal-to-noise ratio 3.4 times higher when implementing the S 2 spectrophone. For the S 2 -based sensor, a linear dependence of the QEPAS signal on ethylene concentration was demonstrated in the 5 ppm-100 ppm range. For a 10 s lock-in integration time, an ethylene minimum detection limit of 10 ppb was calculated.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.