Background: In patients submitted to major pulmonary resection, the postoperative length of stay is mainly influenced by the duration of air leaks and chest tube removal. The measurement of air leaks largely relies on traditional chest drainage systems which are prone to subjective interpretation. Difficulty in differentiating between active air leaks and bubbles due to a pleural space effect may also lead to tentative drain clamping and prolonged time for chest drain removal. New digital systems allow continuous monitoring of air leaks, identifying subtle leakage that may be not visible during daily patient evaluation. Moreover, an objective assessment of air leaks may lead to a reduced interobserver variability and to an optimized timing for chest tube removal. Methods: This study is a prospective randomized, interventional, multicenter trial designed to compare an electronic chest drainage system (Drentech™ Palm Evo) with a traditional system (Drentech™ Compact) in a cohort of patients undergoing pulmonary lobectomy through a standard three-port video-assisted thoracic surgery approach for both benign and malignant disease. The study will enroll 382 patients in three Italian centers. The duration of chest drainage and the length of hospital stay will be evaluated in the two groups. Moreover, the study will evaluate whether the use of a digital chest system compared with a traditional system reduces the interobserver variability. Finally, it will evaluate whether the digital drain system may help in distinguishing an active air leak from a pleural space effect, by the digital assessment of intrapleural differential pressure, and in identifying potential predictors of prolonged air leaks. Discussion: To date, few studies have been performed to evaluate the clinical impact of digital drainage systems. The proposed prospective randomized trial will provide new knowledge to this research area by investigating and comparing the difference between digital and traditional chest drain systems. In particular, the objectives of this project are to evaluate the feasibility and usefulness of digital chest drainages and to provide new tools to identify patients at higher risk of developing prolonged air leaks. Trial registration: ClinicalTrials.gov, NCT03536130. Retrospectively registered on 24 May 2018.

Multicenter randomized study on the comparison between electronic and traditional chest drainage systems

Marulli G.;Larocca V.;Brascia D.;
2019-01-01

Abstract

Background: In patients submitted to major pulmonary resection, the postoperative length of stay is mainly influenced by the duration of air leaks and chest tube removal. The measurement of air leaks largely relies on traditional chest drainage systems which are prone to subjective interpretation. Difficulty in differentiating between active air leaks and bubbles due to a pleural space effect may also lead to tentative drain clamping and prolonged time for chest drain removal. New digital systems allow continuous monitoring of air leaks, identifying subtle leakage that may be not visible during daily patient evaluation. Moreover, an objective assessment of air leaks may lead to a reduced interobserver variability and to an optimized timing for chest tube removal. Methods: This study is a prospective randomized, interventional, multicenter trial designed to compare an electronic chest drainage system (Drentech™ Palm Evo) with a traditional system (Drentech™ Compact) in a cohort of patients undergoing pulmonary lobectomy through a standard three-port video-assisted thoracic surgery approach for both benign and malignant disease. The study will enroll 382 patients in three Italian centers. The duration of chest drainage and the length of hospital stay will be evaluated in the two groups. Moreover, the study will evaluate whether the use of a digital chest system compared with a traditional system reduces the interobserver variability. Finally, it will evaluate whether the digital drain system may help in distinguishing an active air leak from a pleural space effect, by the digital assessment of intrapleural differential pressure, and in identifying potential predictors of prolonged air leaks. Discussion: To date, few studies have been performed to evaluate the clinical impact of digital drainage systems. The proposed prospective randomized trial will provide new knowledge to this research area by investigating and comparing the difference between digital and traditional chest drain systems. In particular, the objectives of this project are to evaluate the feasibility and usefulness of digital chest drainages and to provide new tools to identify patients at higher risk of developing prolonged air leaks. Trial registration: ClinicalTrials.gov, NCT03536130. Retrospectively registered on 24 May 2018.
File in questo prodotto:
File Dimensione Formato  
13063_2019_Article_3811.pdf

accesso aperto

Tipologia: Documento in Versione Editoriale
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 935.55 kB
Formato Adobe PDF
935.55 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11586/259151
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact