Three hundred eighty-five valve prostheses were implanted between 1974 and 1981 in patients with isolated mitral disease: 157 caged-ball valves (156 Starr-Edwards; 1 Smeloff-Cutter) (group A), 107 tilting-disc valves (44 Bjork-Shiley, 49 Sorin, 14 Lillehei-Kaster) (group B), 121 porcine bioprostheses (45 Carpentier-Edwards, 66 Liotta, 10 Hancock) (group C). Perioperative mortality was 9.5% in group A, 11.2% in group B and 6.6% in group C. The follow-up was 86% complete. Actuarial freedom from complications was calculated as follows (linearised rates in brackets) in groups A, B and C, respectively: survival: 47.01% +/- 0.11 (3% patient/yr), 53.37% +/- 0.08 (1.8% patient/yr), 61.24% +/- 0.05 (2.2% patient/yr); thromboembolism: 67.94% +/- 0.09 (1.18% patient/yr); 73.07% +/- 0.06 (1% patient/yr); 97.43% +/- 0.02 (0.02% patient/yr); anticoagulation-related hemorrhage: 84.10% +/- 0.13 (0.18% patient/yr), 97.21% +/- 0.01 (0.12% patient/yr), 100%; prosthetic valve endocarditis: 100% in groups A and B, 95.76% +/- 0.02 (0.18% patient/yr) in group C; valve-related mortality: 87.52% +/- 0.03 (0.75% patient/yr), 87.96% +/- 0.03 (0.56% patient/yr), 82.53% +/- 0.04 (0.93% patient/yr); valve failure: 81.22% +/- 0.07 (0.56% patient/yr), 63.36% +/- 0.1 (1.06% patient/yr), 14.31% +/- 0.05 (4% patient/yr); treatment failure: 78.81% +/- 0.05 (1.12% patient/yr), 76.44% +/- 0.09 (0.62% patient/yr), 80.97% +/- 0.04 (1% patient/yr); all valve-related morbidity and mortality: 40.43% +/- 0.13 (1.93% patient/yr), 57.76% +/- 0.08 (1.43% patient/yr), 14.96% +/- 0.05 (4.18% patient/yr); all valve-related morbidity and mortality at 5 years: 91.97% +/- 0.02 (7.8% patient/yr), 87.06% +/- 0.03 (3.6% patient/yr), 90.27% +/- 0.03 (2.6% patient/yr); at 10 years: 80.4% +/- 0.03 (4.6% patient/yr), 75.91% +/- 0.03 (2.6% patient/yr), 37.44% +/- 0.05 (4.18% patient/yr). The long-term results of the first generation valve prostheses are uniformly unsatisfactory. At least 15 years of follow-up are required for a reliable evaluation as all valve-related morbidity and mortality dramatically increase in the second decade of follow-up.

LONG-TERM FOLLOW-UP OF DIFFERENT MODELS OF MECHANICAL AND BIOLOGICAL MITRAL PROSTHESES

DE LUCA TUPPUTI SCHINOSA, Luigi;
1995-01-01

Abstract

Three hundred eighty-five valve prostheses were implanted between 1974 and 1981 in patients with isolated mitral disease: 157 caged-ball valves (156 Starr-Edwards; 1 Smeloff-Cutter) (group A), 107 tilting-disc valves (44 Bjork-Shiley, 49 Sorin, 14 Lillehei-Kaster) (group B), 121 porcine bioprostheses (45 Carpentier-Edwards, 66 Liotta, 10 Hancock) (group C). Perioperative mortality was 9.5% in group A, 11.2% in group B and 6.6% in group C. The follow-up was 86% complete. Actuarial freedom from complications was calculated as follows (linearised rates in brackets) in groups A, B and C, respectively: survival: 47.01% +/- 0.11 (3% patient/yr), 53.37% +/- 0.08 (1.8% patient/yr), 61.24% +/- 0.05 (2.2% patient/yr); thromboembolism: 67.94% +/- 0.09 (1.18% patient/yr); 73.07% +/- 0.06 (1% patient/yr); 97.43% +/- 0.02 (0.02% patient/yr); anticoagulation-related hemorrhage: 84.10% +/- 0.13 (0.18% patient/yr), 97.21% +/- 0.01 (0.12% patient/yr), 100%; prosthetic valve endocarditis: 100% in groups A and B, 95.76% +/- 0.02 (0.18% patient/yr) in group C; valve-related mortality: 87.52% +/- 0.03 (0.75% patient/yr), 87.96% +/- 0.03 (0.56% patient/yr), 82.53% +/- 0.04 (0.93% patient/yr); valve failure: 81.22% +/- 0.07 (0.56% patient/yr), 63.36% +/- 0.1 (1.06% patient/yr), 14.31% +/- 0.05 (4% patient/yr); treatment failure: 78.81% +/- 0.05 (1.12% patient/yr), 76.44% +/- 0.09 (0.62% patient/yr), 80.97% +/- 0.04 (1% patient/yr); all valve-related morbidity and mortality: 40.43% +/- 0.13 (1.93% patient/yr), 57.76% +/- 0.08 (1.43% patient/yr), 14.96% +/- 0.05 (4.18% patient/yr); all valve-related morbidity and mortality at 5 years: 91.97% +/- 0.02 (7.8% patient/yr), 87.06% +/- 0.03 (3.6% patient/yr), 90.27% +/- 0.03 (2.6% patient/yr); at 10 years: 80.4% +/- 0.03 (4.6% patient/yr), 75.91% +/- 0.03 (2.6% patient/yr), 37.44% +/- 0.05 (4.18% patient/yr). The long-term results of the first generation valve prostheses are uniformly unsatisfactory. At least 15 years of follow-up are required for a reliable evaluation as all valve-related morbidity and mortality dramatically increase in the second decade of follow-up.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11586/105374
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact